AST580 Fully-automatic Cryostat Microtome with Touch Screen
Transform your frozen tissue sectioning workflow with our Fully-automatic Cryostat Microtome equipped with an intuitive touch screen interface. Its advanced automation and precision guarantee swift, reliable sample preparation for precise diagnostics and research endeavors.
Description of AST580 Fully-automatic Microtome Cryostat with Touch Screen
AST580 fully automatic microtome is designed under engineering and ergonomic concepts. All the components are processed by computer numerical control.
Retraction function prevents specimen damage from blade.
Total thickness calculation.
Low temperature refrigeration system with environment friendly refrigerant R404.
Adopt UV sterilize for 30 minutes every time.
Peltier specimen clamp refrigeration can be turn On or Off manually.
Two defrosting methods: manual and automatic.
Emergency stop function: Press the red button to stop running when any emergency happens.
The high prevision unit is enclosed outside the cryostat chamber to prevent it from thermal expansion and contraction, and keeps minimum maintenance requirement.
Artificial intelligent interface is easy to learn and operate.
Touch Screen and display all information including thickness, thickness sum, controlling time, temperature, time on/off and so on.
Sleep Function: after activating this function, the temperature in cryostat chamber can be kept between -1℃ and -10℃; after stopping this function, the temperature in freeze chamber will rise to working temperature within 30 minutes.
Control panel with locking function could avoid faulty operation.
Handwheel locking function.
Wide freezing shelf can load 17 specimen cassettes simultaneously, and two of them are equipped with peltier to meet different temperatures.
Specification of AST580 Fully-automatic Microtome Cryostat with Touch Screen
Environment requirement
Temperature Range
+15℃~+30℃
Air Humidity
≤60%
Working pressure
(86~106)kPa
Nominal Voltage
220~240V AC/100~120V AC
Normal Frequency
50/60 Hz
Power
≤1000VA
Fuse
8A/20A
Safe Classify
Classify I – Type B
Lowest freeze chamber Temperature
-35±2℃
Lowest freeze shelf Temperature
-45±5℃
Lowest temperature of peltier unit on freeze shelf
-55±5℃
Lowest temperature of peltier unit on specimen clamp
-50±5℃(Working time of peltier: 15 minutes)
Electric coarse feed
Slow: 300µm/s±30µm/s
Rapid: 900µm/s±30µm/s
Section thickness range
0 to 100µm0 to 3µm, in 0.5µm increments3 to 10µm, in 1µm increments10 to 20µm, in 2µm increments20 to 100µm, in 5µm increments
Trimming thickness range
0 to 600µm0 to 50µm, in 5µm increments50 to 100µm, in 10µm increments100 to 600µm, in 50µm increments
⊙Retraction
0~80µm, in 5µm increments
⊙Specimen feed
24mm ±0.2mm
⊙Vertical stroke
54mm±1mm
⊙Specimen Max
35×35mm
⊙Repositioning of blade holder base(left-right)
50±1mm
⊙Motorize cutting speed
75~230mm/s ± 10%.
Dimension
Width
715mm
Length
765mm
Height
1230mm
Weight
About 130kgs